Open Ion Source
OIS is the most widely available type of RGA. Cylindrical and axially symmetrical, this kind of ionizer has been around since the early 1950s. The OIS type is usually mounted directly to the vacuum chamber, exposing the filament wire and anode wire cage to the surrounding vacuum chamber, allowing all molecules in the vacuum chamber to move easily through the ion source. With a maximum operating pressure of Torr and a minimum detectable partial pressure as low as Torr when used in tandem with an electron multiplier.
OIS RGAs measure residual gas levels without affecting the gas composition of their vacuum environment, though there are performance limitations which include:
- Outgassing of water from the chamber, from the OIS electrodes and most varieties of 300-series stainless steel used in the surrounding vacuum chamber due to the high temperatures of the hot-cathode source (> 1300°C).
- Electron Stimulated Desorption (ESD) is noted by peaks observed at 12, 16, 19 and 35 u rather than by electron-impact ionization of gaseous species, with the effects similar to outgassing effects. This is frequently counteracted by gold-plating the ionizer which in turn reduces the adsorption of many gases. Using platinum-clad molybdenum ionizers is an alternative.
Read more about this topic: Residual Gas Analyzer
Famous quotes containing the words open and/or source:
“In the greatest confusion there is still an open channel to the soul. It may be difficult to find because by midlife it is overgrown, and some of the wildest thickets that surround it grow out of what we describe as our education. But the channel is always there, and it is our business to keep it open, to have access to the deepest part of ourselves.”
—Saul Bellow (b. 1915)
“Thought is the work of the intellect, reverie is its self-indulgence. To substitute day-dreaming for thought is to confuse a poison with a source of nourishment.”
—Victor Hugo (18021885)