Calibration and Accuracy
Optical Power Meter calibration and accuracy is a contentious issue. The accuracy of most primary reference standards (e.g. Weight, Time, Length, Volt etc.) is known to a high accuracy, typically of the order of 1 part in a billion. However the optical power standards maintained by NIST, are only defined to about one part in a thousand. By the time this accuracy has been further degraded through successive links, instrument calibration accuracy is usually only a few %. The most accurate field optical power meters claim 1% calibration accuracy. Comparatively, this is orders of magnitude less accurate than a typical electrical voltmeter.
Further, the in-use accuracy achieved is usually significantly lower than the claimed calibration accuracy, by the time additional factors are taken into account. In typical field applications, factors may include: ambient temperature, optical connector type, wavelength variations, linearity variations, beam geometry variations, detector saturation.
Therefore, achieving a good level of practical instrument accuracy and linearity is something that requires considerable design skill, and care in manufacturing.
Read more about this topic: Optical Power Meter
Famous quotes containing the word accuracy:
“In everything from athletic ability to popularity to looks, brains, and clothes, children rank themselves against others. At this age [7 and 8], children can tell you with amazing accuracy who has the coolest clothes, who tells the biggest lies, who is the best reader, who runs the fastest, and who is the most popular boy in the third grade.”