Isotope-ratio Mass Spectrometry - Moving Wire IRMS (MW-irMS)

Moving Wire IRMS (MW-irMS)

Moving wire IRMS is useful for analyzing Carbon-13 ratios of compounds in a solution, such as after purification by liquid chromatography. The solution (or outflow from the chromatography) is dried onto a nickel or stainless steel wire. After the residue is deposited on the wire, it enters a furnace where the sample is converted to CO2 and water by combustion. The gas stream finally enters a capillary, is dried, ionized, and analyzed. This process allows a mixture of compounds to be purified and analyzed continuously, which can decrease the analysis time by a factor of four. Moving wire IRMS is quite sensitive, and samples containing as little as 1 nano-mole of Carbon can yield precise (within 1‰) results.

Read more about this topic:  Isotope-ratio Mass Spectrometry

Famous quotes containing the words moving and/or wire:

    One fellow I was dating in medical school ... was a veterinarian and he wanted to get married. I said, but you’re going to be moving to Minneapolis, and he said, oh, you can quit and I’ll take care of you. I said, “Go.”
    Sylvia Beckman (b. c. 1931)

    Constantly risking absurdity and death whenever he performs above the heads of his audience the poet like an acrobat climbs on rime to a high wire of his own making.
    Lawrence Ferlinghetti (b. 1919)