Typical FCS Setup
The typical FCS setup consists of a laser line (wavelengths ranging typically from 405–633 nm (cw), and from 690–1100 nm (pulsed)), which is reflected into a microscope objective by a dichroic mirror. The laser beam is focused in the sample, which contains fluorescent particles (molecules) in such high dilution, that only a few are within the focal spot (usually 1–100 molecules in one fL). When the particles cross the focal volume, they fluoresce. This light is collected by the same objective and, because it is red-shifted with respect to the excitation light it passes the dichroic mirror reaching a detector, typically a photomultiplier tube or avalanche photodiode detector. The resulting electronic signal can be stored either directly as an intensity versus time trace to be analyzed at a later point, or computed to generate the autocorrelation directly (which requires special acquisition cards). The FCS curve by itself only represents a time-spectrum. Conclusions on physical phenomena have to be extracted from there with appropriate models. The parameters of interest are found after fitting the autocorrelation curve to modeled functional forms.
Read more about this topic: Fluorescence Correlation Spectroscopy
Famous quotes containing the word typical:
“The books may say that nine-month-olds crawl, say their first words, and are afraid of strangers. Your exuberantly concrete and special nine-month-old hasnt read them. She may be walking already, not saying a word and smiling gleefully at every stranger she sees. . . . You can support her best by helping her learn what shes trying to learn, not what the books say a typical child ought to be learning.”
—Amy Laura Dombro (20th century)