Coherent Anti-Stokes Raman Spectroscopy - Comparison To Raman Spectroscopy

Comparison To Raman Spectroscopy

CARS is often compared to Raman spectroscopy as both techniques probe the same Raman active modes. Raman can be done using a single continuous wave (CW) laser whereas CARS requires (generally) two pulsed laser sources. The Raman signal is detected on the red side of the incoming radiation where it might have to compete with other fluorescent processes. The CARS signal is detected on the blue side, which is free from fluorescence, but it comes with a non-resonant contribution. The differences between the signals from Raman and CARS (there are many variants of both techniques) stems largely from the fact that Raman relies on a spontaneous transition whereas CARS relies on a coherently driven transition. The total Raman signal collected from a sample is the incoherent addition of the signal from individual molecules. It is therefore linear in the concentration of those molecules and the signal is emitted in all directions. The total CARS signal comes from a coherent addition of the signal from individual molecules. For the coherent addition to be additive, phase-matching must be fulfilled. For tight focusing conditions this is generally not a restriction. Once phase-matching is fulfilled the signal amplitude grows linear with distance so that the power grows quadratically. This signal forms a collimated beam that is therefore easily collected. The fact that the CARS signal is quadratic in the distance makes it quadratic with respect to the concentration and therefore especially sensitive to the majority constituent. The total CARS signal also contains an inherent non-resonant background. This non-resonant signal can be considered as the result of (several) far off-resonance transitions that also add coherently. The resonant amplitude contains a phase shift of π radians over the resonance whereas the non-resonant part does not. The spectral line shape of the CARS intensity therefore resembles a Fano profile which is shifted with respect to the Raman signal. To compare the spectra from multi-component compounds, the (resonant) CARS spectral amplitude should be compared to the Raman spectral intensity.

Theoretically Raman spectroscopy and CARS spectroscopy are equally sensitive as they use the same molecular transitions. However, given the limits on input power (damage threshold) and detector noise (integration time), the signal from a single transition can be collected much faster in practical situation (a factor of 105) using CARS. Imaging of known substances (known spectra) is therefore often done using CARS. Given the fact that CARS is a higher order nonlinear process, the CARS signal from a single molecule is larger than the Raman signal from a single molecule for a sufficiently high driving intensity. However at very low concentrations, the advantages of the coherent addition for CARS signal reduces and the presence of the incoherent background becomes an increasing problem.

Since CARS is such a nonlinear process there are not really any 'typical' experimental numbers. One example is given below under the explicit warning that just changing the pulse duration by one order of magnitude changes the CARS signal by three orders of magnitude. The comparison should only be used as an indication of the order of magnitude of the signals. 200 mW average power input (CW for the Raman), in a 0.9NA objective with a center wavelength around 800 nm, constitutes a power density of 26 MW/cm,2 (focus length = 1.5 micrometre, focus volume = 1.16 micrometre3, photon energy = 2.31 10−19 J or 1.44 eV). The Raman cross section for the vibration of the aromatic ring in Toluene around 1000 cm−1 is on the order of 10−29cm2/molecule·steradian. Therefore the Raman signal is around 26 10−22 W/molecule·steradian or 3.3 10−21 W/molecule (over 4π steradians). That is 0.014 photon/sec·molecule. The density of Toluene = 0.8668 103 kg/m3, Molecular mass = 92.14 10−3 kg/mol. Therefore the focal volume (~1 cubic micrometre) contains 6 109 molecules. Those molecules together generate a Raman signal in the order of 2 10−11 W (20pW) or roughly one hundred million photons/sec (over 4π steradians). A CARS experiment with similar parameters (150 mW at 1064 nm, 200 mW at 803.5 nm, 15ps pulses at 80 Mhz repetition frequency, same objective lens) yields roughly 17.5 10−6 W (on the 3000 cm−1 line, which has 1/3 of the strength and roughly 3 times the width). This CARS power is roughly 106 higher than the Raman but since there are 6 109 molecules, the signal per molecule from CARS is only 4 10−25 W/molecule·s or 1.7 10−6 photons/molecule·s. If we allow two factors of three (line strength and line width) then the spontaneous Raman signal per molecule still exceeds the CARS per molecule by more than two orders of magnitude. The coherent addition of the CARS signal from the molecules however yields a total signal that is much higher than the Raman.

The sensitivity in many CARS experiments is not limited by the detection of CARS photons but rather by the distinction between the resonant and non-resonant part of the CARS signal.

Read more about this topic:  Coherent Anti-Stokes Raman Spectroscopy

Famous quotes containing the word comparison:

    The comparison between Coleridge and Johnson is obvious in so far as each held sway chiefly by the power of his tongue. The difference between their methods is so marked that it is tempting, but also unnecessary, to judge one to be inferior to the other. Johnson was robust, combative, and concrete; Coleridge was the opposite. The contrast was perhaps in his mind when he said of Johnson: “his bow-wow manner must have had a good deal to do with the effect produced.”