Total Ionic Strength Adjustment Buffer - Theory

Theory

TISAB is very commonly applied to fluoride ion analysis such as in fluoride ion selective electrodes.

There are four main constituents to TISAB, namely CDTA (cyclohexylenedinitrilotetraacetate), sodium hydroxide, sodium chloride and acetic acid (ethanoic acid), which are all dissolved in deionised water. Hence, TISAB having a density ~1.0 kg/L. Each constituent plays an important role to control the ionic strength and pH of the analyte solution, which may otherwise cause error and inaccuracy.

The activity of a substance in solution depends on the product of its concentration and the activity coefficient in that solution. The activity coefficient depends on the ionic strength of the solution in which the potentiometric measurements are made. This can be calculated for dilute solutions using the Debye-Hückel equation; for more concentrated solutions other approximations must be used. In most cases, the analyst's goal is simply to make sure that the activity coefficient is constant across a set of solutions, with the assumption that no significant ion pairing exists in the solutions.

Example: An ion-selective electrode might be calibrated using dilute solutions of the analyte in distilled water. If this calibration is used to calculate the concentration of the analyte in sea water (high ionic strength), significant error is introduced by the difference between the activity of the analyte in the dilute solutions and the concentrated sample. This can be avoided by adding a small amount of ionic-strength buffer to the standards, so that the activity coefficients match more closely.

Adding a TISAB buffer to increase the ionic strength of the solution helps to "fix" the ionic strength at a stable level, making a linear correlation between the logarithm of the concentration of analyte and the measured voltage. By also adding the TISAB buffer to the samples from which the potentiometric equipment are calibrated, the linear correlation can be used to calculate the concentration of analyte in the solution.

where is measured voltage, is the gas constant, the temperature measured in kelvins, is the Faraday constant and the charge of the analyte. is the concentration of analyte.

TISAB buffers often include chelators which bind ions that could otherwise interfere with the analyte.

Read more about this topic:  Total Ionic Strength Adjustment Buffer

Famous quotes containing the word theory:

    There is in him, hidden deep-down, a great instinctive artist, and hence the makings of an aristocrat. In his muddled way, held back by the manacles of his race and time, and his steps made uncertain by a guiding theory which too often eludes his own comprehension, he yet manages to produce works of unquestionable beauty and authority, and to interpret life in a manner that is poignant and illuminating.
    —H.L. (Henry Lewis)

    In the theory of gender I began from zero. There is no masculine power or privilege I did not covet. But slowly, step by step, decade by decade, I was forced to acknowledge that even a woman of abnormal will cannot escape her hormonal identity.
    Camille Paglia (b. 1947)

    every subjective phenomenon is essentially connected with a single point of view, and it seems inevitable that an objective, physical theory will abandon that point of view.
    Thomas Nagel (b. 1938)