Moving Wire IRMS (MW-irMS)
Moving wire IRMS is useful for analyzing Carbon-13 ratios of compounds in a solution, such as after purification by liquid chromatography. The solution (or outflow from the chromatography) is dried onto a nickel or stainless steel wire. After the residue is deposited on the wire, it enters a furnace where the sample is converted to CO2 and water by combustion. The gas stream finally enters a capillary, is dried, ionized, and analyzed. This process allows a mixture of compounds to be purified and analyzed continuously, which can decrease the analysis time by a factor of four. Moving wire IRMS is quite sensitive, and samples containing as little as 1 nano-mole of Carbon can yield precise (within 1‰) results.
Read more about this topic: Isotope-ratio Mass Spectrometry