Focused Ion Beam - Principle

Principle

Focused ion beam (FIB) systems have been produced commercially for approximately twenty years, primarily for large semiconductor manufacturers. FIB systems operate in a similar fashion to a scanning electron microscope (SEM) except, rather than a beam of electrons and as the name implies, FIB systems use a finely focused beam of ions (usually gallium) that can be operated at low beam currents for imaging or high beam currents for site specific sputtering or milling.

As the diagram on the right shows, the gallium (Ga+) primary ion beam hits the sample surface and sputters a small amount of material, which leaves the surface as either secondary ions (i+ or i-) or neutral atoms (n0). The primary beam also produces secondary electrons (e–). As the primary beam rasters on the sample surface, the signal from the sputtered ions or secondary electrons is collected to form an image.

At low primary beam currents, very little material is sputtered and modern FIB systems can easily achieve 5 nm imaging resolution (imaging resolution with Ga ions is limited to ~5 nm by sputtering and detector efficiency). At higher primary currents, a great deal of material can be removed by sputtering, allowing precision milling of the specimen down to a sub micrometre or even a nano scale.

If the sample is non-conductive, a low energy electron flood gun can be used to provide charge neutralization. In this manner, by imaging with positive secondary ions using the positive primary ion beam, even highly insulating samples may be imaged and milled without a conducting surface coating, as would be required in a SEM.

Until recently, the overwhelming usage of FIB has been in the semiconductor industry. Such applications as defect analysis, circuit modification, mask repair and transmission electron microscope sample preparation of site specific locations on integrated circuits have become commonplace procedures. The latest FIB systems have high resolution imaging capability; this capability coupled with in situ sectioning has eliminated the need, in many cases, to examine FIB sectioned specimens in a separate SEM instrument. SEM imaging is still required for the highest resolution imaging and to prevent damage to sensitive samples. However, the combination of SEM and FIB columns onto the same chamber enables the benefits of both to be utilized.

Read more about this topic:  Focused Ion Beam

Famous quotes containing the word principle:

    A little grit in the eye destroyeth the sight of the very heavens, and a little malice or envy a world of joys. One wry principle in the mind is of infinite consequence.
    Thomas Traherne (1636–1674)

    Well, you Yankees and your holy principle about savin’ the Union. You’re plunderin’ pirates that’s what. Well, you think there’s no Confederate army where you’re goin’. You think our boys are asleep down here. Well, they’ll catch up to you and they’ll cut you to pieces you, you nameless, fatherless scum. I wish I could be there to see it.
    John Lee Mahin (1902–1984)

    In some things, we Americans leave to other countries the carrying out of the principle that stands at the head of our Declaration of Independence.
    Herman Melville (1819–1891)