Quadrupole Mass Analyzer - Principle of Operation

Principle of Operation

The quadrupole consists of four parallel metal rods. Each opposing rod pair is connected together electrically, and a radio frequency (RF) voltage is applied between one pair of rods and the other. A direct current voltage is then superimposed on the RF voltage. Ions travel down the quadrupole between the rods. Only ions of a certain mass-to-charge ratio m/z will reach the detector for a given ratio of voltages: other ions have unstable trajectories and will collide with the rods. This permits selection of an ion with a particular m/z or allows the operator to scan for a range of m/z-values by continuously varying the applied voltage. Mathematically this can be modeled with the help of the Hill differential equation.

Ideally the rods are hyperbolic. Circular rods with a specific ratio of rod diameter-to-spacing provide an easier-to-manufacture adequate approximation to hyperbolas. Small variations in the ratio have large effects on resolution and peak shape. Different manufacturers choose slightly different ratios to fine-tune operating characteristics in context of anticipated application requirements. In recent decades some manufacturers have produced quadrupole mass spectrometers with true hyperbolic rods.

Read more about this topic:  Quadrupole Mass Analyzer

Famous quotes containing the words principle and/or operation:

    It is funny that men who are supposed to be scientific cannot get themselves to realise the basic principle of physics, that action and reaction are equal and opposite, that when you persecute people you always rouse them to be strong and stronger.
    Gertrude Stein (1874–1946)

    It is critical vision alone which can mitigate the unimpeded operation of the automatic.
    Marshall McLuhan (1911–1980)