Multimeter - Accuracy

Accuracy

Digital multimeters generally take measurements with accuracy superior to their analog counterparts. Standard analog multimeters measure with typically ±3% accuracy, though instruments of higher accuracy are made. Standard portable digital multimeters are specified to have an accuracy of typically 0.5% on the DC voltage ranges. Mainstream bench-top multimeters are available with specified accuracy of better than ±0.01%. Laboratory grade instruments can have accuracies of a few parts per million.

Accuracy figures need to be interpreted with care. The accuracy of an analog instrument usually refers to full-scale deflection; a measurement of 30V on the 100V scale of a 3% meter is subject to an error of 3V, 10% of the reading. Digital meters usually specify accuracy as a percentage of reading plus a percentage of full-scale value, sometimes expressed in counts rather than percentage terms.

Quoted accuracy is specified as being that of the lower millivolt (mV) DC range, and is known as the "basic DC volts accuracy" figure. Higher DC voltage ranges, current, resistance, AC and other ranges will usually have a lower accuracy than the basic DC volts figure. AC measurements only meet specified accuracy within a specified range of frequencies.

Manufacturers can provide calibration services so that new meters may be purchased with a certificate of calibration indicating the meter has been adjusted to standards traceable to, for example, the US National Institute of Standards and Technology (NIST), or other national standards organization.

Test equipment tends to drift out of calibration over time, and the specified accuracy cannot be relied upon indefinitely. For more expensive equipment, manufacturers and third parties provide calibration services so that older equipment may be recalibrated and recertified. The cost of such services is disproportionate for inexpensive equipment; however extreme accuracy is not required for most routine testing. Multimeters used for critical measurements may be part of a metrology program to assure calibration.

Some instrument assume sine waveform for measurements but for distorted wave forms a true RMS converter (TrueRMS) may be needed for correct RMS calculation.

Read more about this topic:  Multimeter

Famous quotes containing the word accuracy:

    As for farming, I am convinced that my genius dates from an older era than the agricultural. I would at least strike my spade into the earth with such careless freedom but accuracy as the woodpecker his bill into a tree.
    Henry David Thoreau (1817–1862)

    The child who has been taught to make an accurate elevation, plan, and section of a pint pot has had an admirable training in accuracy of eye and hand.
    Thomas Henry Huxley (1825–95)

    My attachment has neither the blindness of the beginning, nor the microscopic accuracy of the close of such liaisons.
    George Gordon Noel Byron (1788–1824)